CVE-2026-7251
Description
Eppendorf BioFlo 320Β is vulnerable to due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.
Predictions
Heuristic predictions, AS-IS, for prioritization only.
Mitigations
No mitigations published for this CVE yet.
The vendor-content worker queues fetches as references arrive (check back in a few minutes). Or β if you've already worked around this in production β publish your fix to the community-verified tier.
β Propose a mitigation on Community β Mitigations published via the community go through AI scoring + 2 human reviewers + 7-day silent objection window before landing here withsource_tier=community-verified.
References
CWEs
CWE-259
Community-verified mitigations for this CVE will appear above when contributors publish them.
Verify integrity in audit chain (admin only). AS-IS.